Technology Mad e Ea sy .. . Thermocouple Thermometer Models 60010-00 (Type J) 60010-10 (Type K) 60010-20 (Type T) 68X309905 Rev 1
10 4. BATTERY INSTALLATION AND REPLACEMENT If battery indicator turns on, battery life is approximately 8 to 10 hours. The battery indicator will sta
11 CAUTION: WEAK BATTERIES SHOULD NOT BE LEFT IN THE INSTRUMENT. DEAD BATTERIES CAN LEAK AND CAUSE DAMAGE TO UNIT. DANGER: VOLTAGES PRESENT AT THE
12 5. SELF TEST Press the ON/OFF key. The thermometer performs a self-test and all display digits and indicators, as shown below, should remain on f
13 Thermocouple wiring polarity must be correct. If readings decrease as the temperature increases, the thermocouple wires may be reversed. The red
14
15
16 If a thermocouple is not connected or if the thermocouple is defective, the display will indicate OPEn. Thermocouples are sensitive at the tip or
17 8. HOLD FUNCTON Press the HOLD key to retain the reading on the display. Press the HOLD key again for normal operation. 9. CALIBRATION The therm
18 temperature reading is stable, press the CAL key. The CAL indicator will stop blinking and the reading will be set to 0°C (32°F). The CAL indicato
19 11. FIELD CALIBRATION LOCKOUT AND RE-ENABLE The calibration lockout feature prevents any field calibration changes. The lockout remains in effec
20 13. RE-ENABLE PROCEDURE 1. Turn the thermometer off. 2. Simultaneously press and hold the HOLD and the CAL keys down and momentarily press the ON
21 15. CLEANING WARNING: TO PREVENT IGNITION OFA HAZARDOUS ATMOSPHERE BY ELECTROSTATIC DISCHARGE,CLEAN WITH DAMP CLOTH. Do not clean with abrasives
22 17. ACCESSORIES Replacement Meter and Meter Accessories Item Ordering Code Type J 60010-00 Type K 60010-10 Type T 60010-20 Soft carrying ca
23 General purpose probe (immersion Into liquids), type J EC-TPGLPJ-01M/ 08517-55 Penetration probe (meat, semi-soft Materials), type J EC-TPPEN
24 18. SERVICE WARNING SUBSTITUTION OF COMPONENTS MAY IMPAIR INTRINSIC SAFETY. There are no internal adjustments or user replaceable parts. If “E
25 19. WARRANTY The Manufacturer warrants this product to be free from significant deviations from published specifications for a period of three yea
NOTES
TECHNICAL ASSISTANCE If you have any questions about the use of this product, contact the Manufacturer or authorized seller. Trademarks bearing the
TABLE OF CONTENTS 1. INTRODUCTION 4 2. SAFETY PRECAUTIONS 6 3. SPECIFICATIONS 7 4. BATTERY INSTALLATION AND REPLACEMENT 10 5. SELF TEST 12 6. CONN
4 1. INTRODUCTION This versatile hand-held instrument provides highly accurate temperature measurements. The instrument is designed for easy operat
5 • Low battery warning • Built-in tilt stand for easy hands-free operation • Two blade female ANSI mini-connector input • Operates with a wide se
6 2. SAFETY PRECAUTIONS WARNING: THIS INSTRUMENT IS DESIGNED TO ACCEPT LOW LEVEL SIGNALS SUPPLIED BY STANDARD THERMOCOUPLES. UNDER NO CIRCUMSTANCES S
7 3. SPECIFICATIONS THERMOCOUPLE PROBES Type Temperature range Type J –200°C to 1000°C (–328°F to 1832°F ) Type K –250°C to 1372°C (–418°F to 2
8 Input Protection 50V rms Battery Two AA, 1.5V alkaline ANSI-L40, IEC-LR6. Battery Life: 750 hours continuous, typical. Low battery indication Ba
9 Ingress protection Meets IEC-529 IP-54 for dust and water resistant enclosures. Intrinsic safety This product is energy limited for intrinsicall