SPECIFICATIONS
RTD PROBES
Impedance: 100 ohm @ 0°C (32°F)
Sensitivity (alpha): 0.003850 or 0.003916 ohm/
ohm/°C selectable
Accuracy: ±0.06°F; ±0.03°C at 0.01° resolution
±0.1° at 0.1° resolution
±1° at 1° resolution
LOGGER
Linearization: Conforms to DIN IEC 751, revised
to ITS-90
Input Protection: 10V rms
Display Update: 0.6 seconds per update
Connector: 3-pin circular connector input (mate
to Switchcraft TA3F plug)
Battery
Size: Two AA, 1.5V alkaline ANSI-L40,
IEC-LR6
Life: >300 hours continuous, typical
Display: LCD with 0.4 in high characters main
readout and 0.2 in high characters secondary
displays, 4 digits each display plus various
annunciators.
Temperature/Humidity Range
Operating:
Stated Accuracy: 10°C to 40°C
(50°F to 104°F)
Useful Range: 0°C to 40°C
(32°F to 104°F)
Storage: 10°C to 60°C
(–14°F to 140°F)
Humidity: 10% to 90%
(non-condensing)